Omega Speaker Systems OMEGAFLEX FPU5-MT-220 User Manual
Page 74

Flow rate tests were done with 20°C water at 0 PSI back pressure.
Table 9-2. Average Flow Rates
Tubing Wall
Tubing Size
mL per
Minimum Flow
Maximum Flow
Thickness
Inner Diameter
Revolution
Rate at 600 RPM
Rate at 600 RPM
(mL/Minute)
(mL/Minute)
1
⁄
16
"
1
⁄
32
"
0.05
1
30
1
⁄
16
"
1
⁄
16
"
0.22
3
132
1
⁄
16
"
1
⁄
8
"
0.9
9
540
1
⁄
16
"
3
⁄
16
"
1.9
19
1140
1
⁄
16
"
1
⁄
4
"
3.0
30
1800
1
⁄
16
"
5
⁄
16
"
3.8
38
2280
1.5 mm
1.0 mm
0.08*
1
48
1.5 mm
2.0 mm
0.35*
4
210
1.5 mm
3.0 mm
0.8*
8
482
1.5 mm
4.0 mm
1.43*
15
857
1.5 mm
5.0 mm
2.1*
21
1257
1.5 mm
6.0 mm
2.7*
27
1607
1.5 mm
7.0 mm
3.6*
37
2187
1.5 mm
8.0 mm
3.9*
39
2316
* Metric mL per revolution numbers are rounded off
To determine the chemical compatibility of a particular tubing material, it is
recommended that you test the tubing under actual conditions
Tubing materials that can be used include Vinyl, Viton, Tygon, Silicone, Santoprene,
and Norprene. Up to 68 durometer tubing can be used.
Table 9-3. Variances in Normalized Flow Rate (mL per Revolution)
Tubing Size
Due to
Between Pumps
Due to Different
Inner Diameter
Motor
(due to Mechanical
Tubing Materials
(in.)
Speed
Tolerances)
1
⁄
32
±10%
±25%
±20%
1
⁄
16
±10%
±15%
±15%
1
⁄
8
±10%
±10%
±5%
3
⁄
16
±5%
±10%
±5%
1
⁄
4
±5%
±15%
±5%
5
⁄
16
±15%
±20%
±10%
Specifications
9