Error Message

Problem

Solution

Must STOP first!

You have attempted to calibrate

1.

Press the START/STOP key

or change the direction of

to stop the pump. Verify

pumping while the pump

that the display is in the

is running.

STOPPED Display Mode

.

2.

For information on how to
change the direction of
pumping, refer to Section 5.2.

not applicable

You have attempted to perform

Review the instructions for

a function that you can not

performing the required function

perform at this time.

and proceed.

not available

You have attempted to perform

Review the instructions for

a function that you can not

performing the required function

perform at this time.

and proceed.

OVERCURRENT!

The motor driver board has

1.

The tubing may be causing a

sensed an overcurrent.

jam. Observe how the tubing
is loaded in the pump(s) and,
if necessary, reload it.

2.

Detach the pump(s) and check
for a mechanical jam.
Substitute another pump, if
necessary.

3.

Check to see if the fuse is
blown. Replace if necessary.
Refer to Chapter 9 for fuse rat-
ings and chapter 8 for instruc-
tions on how to replace it.

4.

Make sure you are using tub-
ing of the appropriate wall
thickness for the pump(s).
Refer to Table 9-1, 9-2, or
9-3.

5.

Make sure you are using tub-
ing of the appropriate durom-
eter for your pump(s). Try to
operate the motor with tubing
of 68 or smaller durometer.

Troubleshooting Guide

7

7-4

To recover from an overcurrent,
first turn the power switch off,
and then back on. Then
attempt solutions 1 through 8.