2 troubleshooting – Omron Healthcare M3 User Manual

Page 29

background image

29

5.Handling Errors and Problems

EN

5.2

Troubleshooting

Problem

Cause

Remedy

The reading is

extremely low (or high).

Arm cuff not applied

correctly.

Apply the arm cuff

correctly.
Refer to section 3.2.

Movement or talking

during measurement.

Remain still and do not

talk during

measurement.
Refer to section 3.3.

Clothing is interfering

with the arm cuff.

Remove any clothing

interfering with the arm

cuff.
Refer to section 3.1.

Arm cuff pressure does

not rise.

Is the air jack securely

inserted into the main

unit?

Make sure that the air

jack is connected

securely.
Refer to section 3.2.

Air is leaking from the

arm cuff.

Replace the arm cuff

with a new one.
Refer to Chapter 7.

Arm cuff deflates too

soon.

The arm cuff is loose.

Apply the cuff correctly

so that it is firmly

wrapped around the

arm.
Refer to section 3.2.

Cannot measure or

readings are too low or

too high.

Has the arm cuff been

inflated sufficiently?

Inflate the cuff so that it

is 30 to 40 mmHg above

your previous

measurement result.
Refer to section 3.4.

The unit loses power

during measurement.

The batteries are empty.

Replace the batteries

with new ones.

Nothing happens when

you press the buttons.

The batteries are empty.

Replace the batteries

with new ones.

The batteries have been

inserted incorrectly.

Insert the batteries with

the correct (+/ -) polarity.

7051_E.book Page 29 Friday, July 15, 2005 4:36 PM