5.Handling Errors and Problems

EN

5. Handling Errors and Problems

5.1

Error Messages

Error Display

Cause

Remedy

Cuff is under inflated.

Carefully read and repeat the steps

listed under section 3.4.

Movement during

measurement

Repeat measurement. Remain still

and do not talk during

measurement.
Refer to section 3.3.

Air plug disconnected.

Insert the air plug securely.
Refer to section 3.2.

Arm cuff not applied

correctly.

Apply the arm cuff correctly.
Refer to section 3.2.

Clothing is interfering with

the arm cuff.

Remove any clothing interfering

with the arm cuff.
Refer to section 3.2.

Air is leaking from the arm

cuff.

Replace cuff with new one.
Refer to Chapter 7.

The arm cuff was inflated

above 299 mmHg when

inflating the cuff manually.

Release your finger from the O/I

START button before the pressure

reaches 299 mmHg, when inflating

the cuff manually.
Refer to section 3.4.

Battery power is low.

Replace all four “AA” batteries with

new ones.
Refer to section 2.1.

Blinks or appears

continuously