Operating environment, Storage environment, Shock and vibration – National Instruments Module NI PXI-8250 User Manual
Page 27
Appendix A
Specifications
© National Instruments Corporation
A-3
Operating Environment
Ambient temperature range.................... 0 to 55 °C
(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 low
temperature limit and
MIL-PRF-28800F Class 2 high
temperature limit.)
Relative humidity range ......................... 10% to 90%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Storage Environment
Ambient temperature range.................... –40 to 70 °C
(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)
Relative humidity range ......................... 5% to 95% noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Shock and Vibration
Operating shock ..................................... 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
operating................................................. 5 to 500 Hz, 0.3 g
rms
Non-operating ........................................ 5 to 500 Hz, 2.4 g
rms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.)