Figure 3-4. above-high-level triggering mode, Figure 3-5. inside-region triggering mode, Figure 3-6. high-hysteresis triggering mode – National Instruments PCI-4452 User Manual
Page 25: Figure 3-4, Above-high-level triggering mode -8, Figure 3-5, Inside-region triggering mode -8, Figure 3-6, High-hysteresis triggering mode -8

Chapter 3
Hardware Overview
PCI-4451/4452 User Manual
3-8
©
National Instruments Corporation
In above-high-level triggering mode, the trigger is generated when the
signal value is greater than highValue. LowValue is unused.
Figure 3-4. Above-High-Level Triggering Mode
In inside-region triggering mode, the trigger is generated when the signal
value is between the lowValue and the highValue.
Figure 3-5. Inside-Region Triggering Mode
In high-hysteresis triggering mode, the trigger is generated when the signal
value is greater than highValue, with the hysteresis specified by lowValue.
Figure 3-6. High-Hysteresis Triggering Mode
highValue
Trigger
highValue
Trigger
lowValue
highValue
Trigger
lowValue
User.book Page 8 Tuesday, April 14, 1998 10:20 AM