Chapter 5 Calibration

National Instruments Corporation

5-3

VXI-MIO Series User Manual

module by calling the NI-DAQ or VXI

plug&play

instrument driver

calibration function.

To externally calibrate your module, be sure to use a very accurate

external reference. The reference should be several times more accurate
than the module itself. For example, to calibrate a 12-bit module, the
external reference should be at least

±

0.005% (

±

50 ppm) accurate. To

calibrate a 16-bit module, the external reference should be at least

±

0.001% (

±

10 ppm) accurate.

Other Considerations

The CalDACs adjust the gain error of each analog output channel by

adjusting the value of the reference voltage supplied to that channel.
This calibration mechanism is designed to work only with the internal
10 V reference. Thus, in general, it is not possible to calibrate the
analog output gain error when using an external reference. In this case,
it is advisable to account for the nominal gain error of the analog output
channel either in software or with external hardware. See Appendix A,

Specifications

, for analog output gain error information.