Chapter 4 Signal Connections

National Instruments Corporation

4-15

VXI-MIO Series User Manual

Differential Connections for Ground-Referenced
Signal Sources

Figure 4-4 shows how to connect a ground-referenced signal source to
a channel on a VXI-MIO Series module configured in DIFF input
mode.

Figure 4-4.

Differential Input Connections for Ground-Referenced Signals

With this type of connection, the PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the
signal source and the VXI-MIO Series module ground, shown as V

cm

in Figure 4-4.

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

Other Input Multiplexers

AISENSE

Instrumentation

Amplifier

V

m

Measured

Voltage

AIGND

-

+

-

+

V

s

V

cm

+

-

-

+

I/O Connector

ACH<0..7>

ACH<8..15>

Selected Channel in DIFF Configuration

PGIA