Appendix C Common Questions

VXI-MIO Series User Manual

C-4

National Instruments Corporation

a.

Enable the PFI5 line for output, as follows:

If you are using NI-DAQ, call

Select_Signal(deviceNumber, ND_PFI_5,

ND_OUT_UPDATE, ND_HIGH_TO_LOW)

.

If you are using LabVIEW, invoke the Route Signal VI with

signal name

set to

PFI5

and

signal source

set to

AO Update

.

b.

Set up data acquisition timing so that the timing signal for
A/D conversion comes from PFI5, as follows:

If you are using NI-DAQ, call

Select_Signal(deviceNumber, ND_IN_CONVERT,

ND_PFI_5, ND_HIGH_TO_LOW)

.

If you are using LabVIEW, invoke the AI Clock Config VI
with

clock source code

set to

PFI pin

,

high to low

, and

clock

source string

set to

5

.

c.

Initiate analog input data acquisition, which will start only
when the analog output waveform generation starts.

If you are using NI-DAQ, call

DAQ_Start

with appropriate

parameters.

If you are using LabVIEW, invoke the AI Control VI with

control code

set to

0

(start).

d.

Initiate analog output waveform generation.

If you are using NI-DAQ, call

WFM_Group_Control

with

operation

set to

1

(start).

If you are using LabVIEW, invoke the AO Control VI with

control code

set to

0

(start).

Timing and Digital I/O

12. What types of triggering can be hardware implemented on my

VXI-MIO Series module?

Digital triggering is supported by hardware on every
VXI-MIO Series module. In addition, the VXI-MIO-64E-1 and
VXI-MIO-64XE-10 support analog triggering in hardware.

13. What added functionality does the DAQ-STC make possible in

contrast to the Am9513?

The DAQ-STC incorporates much more than just 10 Am9513-style
counters within one chip. In fact, the DAQ-STC has the complexity
of more than 24 chips. The DAQ-STC makes possible PFI lines,