NI USB-6008/6009 User Guide and Specifications

Safety

Standards

The NI USB-6008/6009 is designed to meet the requirements of the
following standards of safety for electrical equipment for measurement,
control, and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1, CSA 61010-1

Note

For UL and other safety certifications, refer to the product label, or visit

ni.com/certification

, search by model number or product line, and click the

appropriate link in the Certification column.

Voltages

Connect only voltages that are within these limits.

Channel-to-GND

.................................... ±30 V max,

Measurement Category I

Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.

Caution

Do not use this module for connection to signals or for measurements within

Measurement Categories II, III, or IV.

Hazardous Locations

The NI USB-6008/6009 are not certified for use in hazardous locations.

Environmental

The NI USB-6008/6009 device is intended for indoor use only.

Operating temperature
(IEC 60068-2-1 and IEC 60068-2-2) ..... 0 to 55 °C

Operating humidity
(IEC 60068-2-56) ................................... 5 to 95% RH, noncondensing

Maximum altitude .................................. 2,000 m (at 25 °C ambient

temperature)