beautypg.com

Specifications, Input characteristics – National Instruments cFP-RTD-124 User Manual

Page 11

background image

© National Instruments Corp.

11

FP-RTD-124 and cFP-RTD-124

Specifications

These specifications are typical for the range –40 to 70 °C unless
otherwise noted.

Input Characteristics

Number of channels.......................... 8

ADC resolution................................. 16 bits

Type of ADC..................................... Delta-sigma

Input signal ranges (software-selectable by channel)

Temperature................................ 73 to 1123 K

–200 to 850 °C
–328 to 1562 °F

Resistance................................... 0 to 400

Temperature accuracy

Resolution ......................................... 0.016 °C

Resistance accuracy

Offset error ................................. 0.03

Ω typical, 0.1 Ω max

Gain error ................................... 0.06% typical, 0.1% max
Resolution................................... 0.0061

Excitation current.............................. 135 ms pulses of 2 mA every

second

Input noise

........................................±1 bit peak-to-peak

Signal input bandwidth ..................... 3 Hz

Update rate........................................ Each channel is updated

once a second

Error

Measured Value

15 to 35 °C

–40 to 70 °C

Typical

Maximum

Typical

Maximum

–200 to 150 °C

0.15

0.25

0.30

0.60

150 to 850 °C

0.25

0.50

0.80

1.50