beautypg.com

Safety, Maximum voltage, Isolation voltages – National Instruments NI 9265 User Manual

Page 17

background image

© National Instruments Corp.

17

NI 9265 Operating Instructions and Specifications

Safety

Maximum Voltage

1

Connect only voltages that are within the following limits.

Channel-to-COM or V

sup

-to-COM ... ±40 V max

Isolation Voltages

Channel-to-channel........................... No isolation between

channels

Channel-to-earth ground, V

sup

-to-earth ground,

or COM-to-earth ground

Continuous ................................. 250 V

rms

,

Measurement Category II

Withstand.................................... 2,300 V

rms

, verified by a 5 s

dielectric withstand test

Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as

1

The maximum voltage that can be applied between any channel or V

sup

terminal

and a COM terminal without damaging the module or other devices.