NI 9215 with Screw Terminal Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-COM .............................. ±30 V max
Isolation

Channel-to-channel .................... No isolation between

channels

Channel-to-earth ground

Continuous ........................... 250 V

rms

,

Measurement Category II

Withstand ............................. 2,300 V

rms

, verified by a

5 s dielectric withstand test

Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.