NI 9215 Operating Instructions and Specifications

20

ni.com

Input voltage ranges

1

1

Refer to the

Safety Guidelines

section for more information about safe operating

voltages.

Measurement Voltage,

AI+ to AI–

Maximum Voltage

(Signal + Common Mode)

Minimum*

(V)

Typical

(V)

Maximum

(V)

Screw

Terminal

BNC

±10.2

±10.4

±10.6

Each channel
must remain
within
±10.2 V of
common.

All inputs
must remain
within 10.2 V
of the average
AI

inputs.

*

The minimum measurement voltage range is the largest voltage the NI 9215 is

guaranteed to accurately measure.