beautypg.com

Correction mode, 1 changing the target constant – Pentax R-315EX(NX) User Manual

Page 43

background image

43

7. CORRECTION MODE

7.1 Changing the target constant

Changing the Target Constant can be performed only when the Reflector sheet and Prism
Constant settings are “INPUT” in Initial Setting 1.

Example: Prism Constant - 25mm setting

Press [F4] [CORR] in mode B.
(If the instrument is in mode A,
press [F5] [MODE] to enter mode B.)
(SHEET CONST: Reflector sheet constant)

Press the [F5] [SELECT] to enable
the Prism Constant to be changed.

Clear the exiting values by pressing [CLEAR] key. Input 25 by pressing the numeric keys.

Press the [ENT] key to accept the Prism
Constant to -25mm.

Pressing the [ENT] key returns the instrument
to mode A.

To set the Reflector sheet constant to “0” select “0” for “Prism Constant” in “Initial Setting 1”.

To set the Prism constant to “0” or “- 30” select “0” or “- 30” for “Prism Constant” in “Initial
Setting 1”.

When the “Reflector sheet Constant” has been set to “0” in Initial Setting 1 and “Prism
Constant” has been set to “0” or “- 30",“ ” is displayed to the left of “0” or “- 30” on the
correction menu screen. When “ ” is on the screen, the Constant cannot be changed
(by entering a numeric key).

1. PRISM CONST :

-30mm

2. SHEET CONST *

0mm

3. TEMP

* 15°C

4. PRESS

*

1013hPa

5.ppm

* 0

ppm

SELECT

CORRECTION

1. PRISM CONST :

-3 0 mm

2. SHEET CONST *

0mm

3. TEMP

* 15°C

4. PRESS

*

1013hPa

5.ppm

* 0

ppm

CLEAR

CORRECTION

H.angle

H.dst.

V.dst.

MEAS TARGET

0 SET DISP MODE

MODE A

15°C

S 0

92° 30’ 20”

SELECT

CORRECTION

1. PRISM CONST :

-25mm

2. SHEET CONST *

0mm

3. TEMP

* 15°C

4. PRESS

*

1013hPa

5.ppm

* 0

ppm

This manual is related to the following products: