Lake Shore Model 642 Electromagnet Power Supply User’s Manual

5-34

Computer Interface Operation

LIMIT

Limit Output Settings Command

Input:

LIMIT , [term]

Format:

+nn.nnnn, +nn.nnnn

Specifies the maximum output current setting allowed: 0 – 70.1000 A.

Specifies the maximum output current ramp rate setting allowed: 0.0001 – 99.999 A/s.

Remarks:

Sets the upper setting limits for output current, compliance voltage, and output current ramp rate. This
is a software limit that will limit the settings to these maximum values.

LIMIT?

Limit Output Settings Query

Input: LIMIT?

[term]

Returned: , [term]
Format:

+nn.nnnn, +nn.nnnn

(Refer to command for description)

LOCK

Keypad

Lock

Command

Input:

LOCK , [term]

Format: n,nnn

0 = Unlock, 1 = Lock All, 2 = Lock Limits.

Specifies lock-out code. Valid entries are 000

999.

Remarks:

Locks out all front panel entries operations.

Example: LOCK

1,123[term] Enables keypad lock and sets the code to 123.

LOCK?

Keypad

Lock

Query

Input: LOCK?[term]
Returned:
,

[term]

Format:

n,nnn

(Refer to command for description)

MAGWTR

Magnet Water Mode Command

Input: MAGWTR

[term]

Format:

n

0 = Manual Off, 1 = Manual On, 2 = Auto, 3 = Disabled.

Example: MAGWTR

2[term]

Places the magnet water mode to Auto, which will automatically control the

magnet water valve based on the calculated output power.

MAGWTR?

Magnet Water Mode Query

Input: MAGWTR?[term]
Returned:
[term]
Format:

n

(Refer to command for description)

MODE

IEEE Interface Mode Command

Input: MODE

[term]

Format:

n

0 = Local, 1 = Remote, 2 = Remote with local lockout.

Example:

MODE 2[term] – Places the Model 642 into remote mode with local lockout.

MODE?

IEEE Interface Mode Query

Input: MODE?[term]
Returned:
[term]
Format:

n

(Refer to command for description)