Lake Shore Model 642 Electromagnet Power Supply User’s Manual

7-22

Service


CALG?

Gain Calibration Constant Query

Input:

CALG? [term]

Format: nn

0 – 10.

Returned: [term]

Format: ±nnnnnnn

(Refer to command for description)


CALSAVE

Calibration Save Command

Input:

CALSAVE[term]

Remarks:

Saves all CALZ and CALG calibration constants in non-volatile memory.


CALZ

Zero Offset Calibration Constant Command

Input:

CALZ , [term]

Format: nn,

±nnnnnnn

Specifies the item to calibrate. Valid entries are:

0 = Output I Reading
1 = Bias A Reading *
2 = Bias B Reading *
3 = Gnd Diff Reading *
4 = Out Con Reading *
5 = Output V Reading
6 = Boost V Reading*
7 = Ext Prog Input Reading
8 = Temp Reading*
9 = Out Stg V Reading *
10 = Actual Output Current

Zero offset calibration constant value.

Remarks:

Items marked with a * are for internal diagnostic use only and should always be set
to a value of 0 (default).