SYMPTOM

CAUSE/REMEDY

Power is not supplied.

Is the power cord connected? Replace the power cord.
Is there a connection to the main switch?

Replace the main switch: P/N 1A154A084-01 (560 or 750 Series).

Circuit breaker CB1 trips when

main switch is turned to LAMP

position.

There may be a short circuit of the fan motor or transformer T-10.
Replace the defective part (depending on your model) :

P/N 1A166A106-03

P/N 1A669A140-01 (M4)

P/N 1A6699A107-01 (M3)

P/N 1A165F033-06 (M2)

P/N 1A165F039-01 (M5, M6)

P/N 1A165F033-06 (M1)

P/N 1A164A101-01 (T-10)

Xenon lamp will not start (starter

spark is detected when the main

switch is turned to the LAMP

position).

Is the lamp positive wire too close to the metal of the body frame?

Change the position of the lamp wire. It should be at least 15mm from

all metal.
There may be a defective Xenon lamp. This is one indication that the

lamp life has expired.

Replace the lamp (P/N XB-560 or P/N XB-750, depending on your

model).

No starter spark is detected when

the main switch is turned to the

LAMP position.

85V DC may be across the starter or the starter may be defective.

Replace the starter with P/N 1A685A088-01 (560 Series, no housing), P/N

1A159A035-01 (560 Series, black housing) or P/N 1A817S006-01 (750

Series).
70V or less is across the starter, +12V is not present between pins 15 &

14 of IC Q1.

Check the connector: CN12 P/N 1A172A058-03 (560 Series) or CN52

P/N 1A172B070-03 (750 Series).

Dark Xenon lamp.

The lamp current may be under the current of 25 Amps for the 560 Series

or under 37.5 Amps for the 750 Series.

Adjust VR11 of pulse width PCB for the 560 Series or VR1 of pulse width

PCB for the 750 Series.
The lamp may be worn out.

Replace the lamp (P/N XB-560 or P/N XB-750, depending on your

model).

Dark picture or uneven illumination

of image.

The lamp or optics may be improperly positioned.

Adjust the lamp position and check alignment of reflectors and the

reflector holders.

The Xenon lamp may be defective.

Replace the lamp (P/N XB-560 or P/N XB-750, depending on your

model).
NOTE: DON’T BREAK THE PAINT ON THE REFLECTOR MOUNT

SCREWS TO CHECK IF THEY ARE LOOSE. A COMMON PROBLEM IS

THAT THE REFLECTOR HAS BECOME LOOSE FROM REPEATED

SHIPPING.

The biggest problem that will occur with these units is failure to perform preventative maintenance. After a unit

has been shipped a number of times it is a good idea to check the electrical connections and to look for loose

screws. If loose screws are found and tightened, you will save many problems later.