5 MAINTENANCE AND TROUBLESHOOTING
31
5 Maintenance and Troubleshooting
5.1
Cleaning the System
1.
Remove the bedding.
2.
If necessary, inflate the mattress.
3.
Ensure that the control unit is off.
4.
Unplug the power cord from the wall outlet.
5.
Ensure that the underside of the mattress is clear of all sharp objects.
6.
Examine the surface of the control unit and mattress assembly components for visible blood or body fluids.
7.
Perform one of the following:
•
If blood is present, decontaminate the product.
i.
Remove all visible soil with disposable paper towels.
ii.
Scrub the area with freshly prepared effective phenolic detergent disinfectant solution.
ƽ
WARNING
Before cleaning or disassembling the MA90 series, check the underside of the mattress folds for sharp objects such as
scissors, needles, etc. These objects should be removed and discarded before proceeding with further cleaning or
disassembly; otherwise injury or damage to the product may occur.
Because of potential risk of infectious exposure, cleaning with the patient on the bed is not recommended.
All equipment should be inspected. Any item that is visibly soiled with the patient's blood or other body fluids should be
properly cleaned or removed.
Laundry workers should treat all soiled bedding as if it were contaminated with pathogenic microorganisms.