4 OPERATION
29
4.11 Preparing for CPR Procedure
1.
Perform one of the following:
•
MA90Z Power Unit - Press and hold the tabs on the hose connector (Detail “B”).
•
MA95Z Power Unit - Lift the connector cover (Detail “A”).
2.
Pull the hose connector (Details “B” and “C”) from the control unit connectors.
3.
MA90Z Only - Disconnect the RED CPR connector located on the hose (Detail “D”).
For this procedure, refer to FIGURE 5 on page 30.
MA95Z Only - If the hose connector is difficult to release, pull the hose connector down and then away.