4 OPERATION

Part No 1148139

25

Invacare

®

MA90Z Series Mattresses

Max Inflate Button (MA95Z Power Unit)

1.

Press the Max Inflate button (

) to select the max inflate mode and inflate the mattress rapidly to maximum firmness (pressurized to 35 ± 6

mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 15 minutes, the Max Inflate mode
deactivates and the control unit defaults to the previous setting.

2.

Perform one of the following:

MA90Z Power Unit - Press the Max Inflate button again to manually disengage Max Inflate mode.

MA95Z Power Unit - Select a different setting to exit Max Inflate mode, or press the Max Inflate button again.

Lock/Alarm Silence Button

1.

Press the Lock/Alarm Silence button (

) to select one of the following modes:

Lock Mode - Pressing this button locks out all control unit functions, including the On button, to prevent any tampering with the settings.
Press the button for approximately 3-5 seconds for the Lock LED to activate.

Alarm Silence mode - Pressing this button silences the alarm that sounds in the event of power failure or when the hose is disconnected
from the control unit.

It takes 25-60 seconds for the MA95Z mattress to inflate fully (inflation time depends on size of mattress).
It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or
patient cleaning.

This will deactivate the Max Inflate LED.

The Lock LED illuminates when in this mode.

The Alarm LED illuminates when in this mode.

This manual is related to the following products: