SECTION 6—MAINTENANCE

Part No 1143482

25

Perfecto

2

Series

FIGURE 6.2 Preventive Maintenance Record

ON EACH INSPECTION

MODEL NO.

I RC

SERIAL NO.

PREVENTIVE MAINTENANCE RECORD

Record Date of Service

Record Elapsed Hours On Hour Meter

Clean Cabinet Filter(s)

Check Prescribed L/min. Flow Rate

EVERY

4,380

H

OURS

Check Oxygen Concentration

(UNITS WITHOUT

SensO

2

)

Clean/Replace Cabinet Filter(s)

Check Outlet HEPA Filter*

Check Compressor Inlet Filter*

(

Check Power Loss Alarm

AS REQUIRED

Clean Heat Exchanger

Replace Exhaust Muffler

Rebuild Top End of Compressor

*NOTE:

Refer to Preventive Maintenance section of Service Manual.

NOTE:

4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.

26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.

Check Oxygen Concentration

UNITS WITH SensO

2

)

DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS

E

V

ERY 2

6

,280

HOUR

S

OR 3 YEARS, WHICHEVER COMES FIRST