SECTION 4—TYPICAL PRODUCT PARAMETERS
12
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input:
3 A
Sound Level:
43 ± 2 dBA Average for Perfecto
2
Models and
Perfecto
2
V Model
40 ± 2 dBA Average for Perfecto
2
W Models
Altitude:
All Perfecto
2
Models: Up to 8,000 ft (2438
meters) above sea level without degredation of
concentration levels. From 8,000 ft to 13,129
(4000 meters) below 90% efficiency.
All Perfecto
2
Models while filling a
HomeFill System: Up to 6,000 ft (1828 meters)
above sea level without degradation of
concentration levels. From 6,000 ft (1828 meters)
to 13,129 ft (4000 meters) below 90% efficiency.
*Oxygen Output
Concentration Levels:
87% to 95.6% for flows ranging at 0.5 to 5 L/min
Maximum Outlet Pres-
sure:
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range:
0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Low Flow Alarm:
0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories
Connected).
Average Power
Consumption:
Perfecto
2
and Perfecto
2
W Models: 280 W
(Operating @ 3 L/min)
Perfecto
2
V Model: 325 W
Pressure Relief Mecha-
nism Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)