SECTION 4—TYPICAL PRODUCT PARAMETERS

Perfecto

2

Series

12

Part No 1143482

SECTION 4—TYPICAL PRODUCT
PARAMETERS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60 Hz

Rated Current Input:

3 A

Sound Level:

43 ± 2 dBA Average for Perfecto

2

Models and

Perfecto

2

V Model

40 ± 2 dBA Average for Perfecto

2

W Models

Altitude:

All Perfecto

2

Models: Up to 8,000 ft (2438

meters) above sea level without degredation of
concentration levels. From 8,000 ft to 13,129
(4000 meters) below 90% efficiency.

All Perfecto

2

Models while filling a

HomeFill System: Up to 6,000 ft (1828 meters)
above sea level without degradation of
concentration levels. From 6,000 ft (1828 meters)
to 13,129 ft (4000 meters) below 90% efficiency.

*Oxygen Output
Concentration Levels:

87% to 95.6% for flows ranging at 0.5 to 5 L/min

Maximum Outlet Pres-
sure:

5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)

Flow Range:

0.5 to 5 L/min
For flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.

Low Flow Alarm:

0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories
Connected).

Average Power
Consumption:

Perfecto

2

and Perfecto

2

W Models: 280 W

(Operating @ 3 L/min)

Perfecto

2

V Model: 325 W

Pressure Relief Mecha-
nism Operational at:

35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)