SECTION 5—OPERATING INSTRUCTIONS

Part No 1143482

21

Perfecto

2

Series

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm‐up 
time (approximately 30 min.) while waiting for the O

2

 purity to 

reach maximum.

Front Panel Indicators - Units without
SensO

2

NOTE: For this procedure, refer to FIGURE 5.7.

FIGURE 5.7 Front Panel Indicators - Units without SensO2

LABEL

SYMBOL

O

2

PURITY

INDICATOR LIGHTS (LED)

SYSTEM OKAY
O

2

over 85% (± 2%)

GREEN Indicator Light

O

2

Between 73% (± 3%)

to 85% (±2%)

YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician. Have a
back-up supply of oxygen ready.

SYSTEM FAILURE
O

2

Below 73% (±3%)

RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down. Call a qualified technician.
Switch to a back-up supply of oxy-
gen immediately.

O

2

I/O

Auto Shut Down

Use Backup

Normal

RED

GREEN

Call Supplier