beautypg.com

GE E4H-CE Series User Manual

Page 36

background image

TROUBLESHOOTING

SYMPTOM

Electrical machine shut

down

(continued)

Conductivity drifts or

changes after calibration

Display is blank

pH reading is off by more

than 1pH unit

POSSIBLE CAUSES

Motor starter overloaded,

heater tripped.

Calibration may have been

done before the reading stabi-

lized

The stable line may contain

electrical noise

Turbine meter not spinning

Monitor not powered

Probe calibrated using span or

the pH probe is bad

REMEDIES

Turn the switch OFF; let the

heater(s) cool.

When calibrating, wait at

least 15 seconds to 1-1/2

minutes for reading to stabi-

lize.

To reduce electrical noise,

use grounded metal fittings

on the inlet and outlet of

the monitor plumbing.

Check meter for spin.

Open the front panel.

Look at the yellow LEDs

on the rear power board.

Are they ON? If yes, check

ribbon cable. Is it properly

seated? If no, does the unit

have power?

Re-initialize calibration.

Recalibrate using zero

only. Replace pH sensor.

31

This manual is related to the following products: