Main out, Aux out, Ext trig – Fluke DDS Function Generator 271 User Manual
Page 18: Rear panel connectors
![background image](/manuals/105156/18/background.png)
271
Getting Started Manual
Front Panel Inputs and Outputs
MAIN OUT
MAIN OUT
is the 50 Ω or
600 Ω output from the main
generator. It will provide up
to 20 V p-p into a high-
impedance load or 10 V p-p
into a matched 50 Ω or
600 Ω load. It can tolerate a
short circuit for 60 seconds.
shx0005f.gif
AUX OUT
The AUX OUT socket
provides a two-level TTL
and CMOS compatible
signal, synchronous with
the signal from the
MAIN OUT
socket.
shx0006f.gif
The symmetry of the AUX OUT signal is the same as that
set for the main output but you can adjust the phase
relationship between MAIN OUT and AUX OUT using the
PHASE
setting in the TRIG menu.
AUX OUT
logic levels are nominally 0 V and 5 V from
typically 50 Ω. AUX OUT will withstand a short-circuit.
EXT TRIG
EXT TRIG
is the external
trigger input for the non-
continuous modes (trigger,
gate, sweep, FSK and hop)
and for synchronizing the
generator as a slave to an
external master generator.
shx0012f.gif
8