Arbitrary waveforms – Fluke DDS Function Generator 271 User Manual
Page 33
Getting Started
Using the Instrument
Saving and Recalling Settings
Having done all this work it would be wise to save the
settings for future use. You can do this simply by pressing
the STORE key, followed by a number in the range 1 to 9
and the ENTER key. Store 0 is reserved for the factory
default settings. You can recall the set-up at any time
using the RECALL key.
Arbitrary Waveforms
Arbitrary waveforms are best defined on a PC and
downloaded to the instrument via the RS232 or GPIB
interface. Each is identified by a name with up to 16
characters.
Several commonly-used waveforms are available from the
internal read-only memory (ROM) and can be recalled by
pressing the EDIT then the ARB keys.
The display will look something like this:
RECALL ARB No: 14
SINX/X
ENTER TO EXECUTE
Or possibly like this:
RECALL ARB No: 03
ARB 03 IS EMPTY
ENTER TO EXECUTE
The ROM is used for the factory-preset waveforms 06 to
21 inclusive, and the internal non-volatile RAM is used for
23