C.27 status:preset command, Figure c-5. using status commands and queries, C.28 status:questionable[:event]? query – KEPCO TMA VXI-27 User Manual
Page 57: C.27, C.28, Stat:pres, Stat:ques

C-8
TMA VXI-27 101602
C.27 STATus:PRESet COMMAND
STAT:PRES
Syntax:
Short Form: STAT:PRES
Long Form: STATus:PRESet
Description: Disables reporting of all status events. This command sets all bits of the Operation Condition
(Table C-2) and Questionable Condition Registers to 0, preventing all status events from being
reported. (See example, Figure C-5.)
FIGURE C-5. USING STATUS COMMANDS AND QUERIES
C.28 STATus:QUEStionable[:EVENt]? QUERY
STAT:QUES?
Syntax:
Short Form: STAT:QUES[EVEN]?
Long Form: STATus:QUEStionable[EVENT]?
Return Value:
Description: Indicates questionable events that occurred since previous STAT:QUES? query. Returns the
value of the Questionable Event register (see Table C-3). The Questionable Event register is a
read-only register which holds (latches) all events. Reading the Questionable Event register clears it.
NOTE: The selected power supply is assumed to be operating in cV (constant voltage) mode.
STAT:OPER:ENAB 1056
Mask enabled for CC, WTG and bits.
STAT:OPER:ENAB?
Returns 1056 (32 + 1024) (CC, WTG bits set).
STAT:QUES:ENAB 3
Mask enabled for OV and OC bits (1 + 2).
STAT:QUES:ENAB?
Returns 3 (1 + 2) indicating OV and OC bits are enabled.
STAT:PRES
Operation Condition and Questionable Condition registers are
reset.
INIT:CONT ON
Continuous triggers enabled.
STAT:OPER:COND?
Power supply returns 288 (256 + 32) to indicate that power
supply is constant voltage mode and Wait For Trigger is true.
STAT:OPER?
Returns 1057, e.g., indicating that since the last reading of the
Operation Event Register the power supply has entered
Constant Current mode, the Wait Trigger was set.
STAT:OPER?
Returns 0 indicating no changes since previous reading of the
Operation Event register.
STAT:QUES?
Returns 0 (no questionable conditions occurred since previous
reading
--- OVERCURRENT CONDITION OCCURS
STAT:QUES?
Returns 2 (overcurrent protection tripped since the last
STAT:QUES? query).
STAT:QUES:COND?
Returns 2, (Power supply still in overcurrent protection state).
STAT:QUES?
Returns 0, (Register cleared by previous STAT:QUES?).
STAT:QUES:COND?
Returns 2, (Power supply still in overcurrent protection state).
SYST:ERR?
Power supply returns 0,“No error” message.
INST:SEL 1
Select instrument 1
STAT:QUES?
TMA responds 4 indicating instrument 1 in current mode
INST:SEL 2
Select instrument 2
STAT:QUES?
TMA responds 8 indicating instrument 2 in voltage mode
STAT:QUES:INST:ISUM1?
TMA responds 4 indicating instrument 1 in current mode
STAT:QUES:INST:ISUM2?
TMA responds 8 indicating instrument 2 in voltage mode