C.12 output[:state] command, C.13 output[:state] query, C.12 – KEPCO TMA VXI-27 User Manual
Page 53: C.13, C.14, C.15, C.12, c.13, C.14, c.15, Outp, Curr

C-4
TMA VXI-27 101602
NOTE: The SCPI convention for this command allows the controller to establish the range and accu-
racy of the measurement if nn,nn is added after the question mark; the power supply accepts this for-
mat but sets the command warning bit (13) in the status questionable register and ignores the extra
characters. (See example, Figure C-3.)
C.12 OUTPut[:STATe]
COMMAND
OUTP
Syntax:
Short Form: OUTP[:STAT]
Long Form: OUTPut[:STATe]
OUTP
OUTP
Description: Enables or disables the power supply output. Upon power up the output is disabled (OUTP OFF),
except for BOP power supplies which power up with the output enabled (OUTP ON). When OUTP
OFF is executed, the programmed values of voltage and current are saved, then voltage and current
are programmed to 0. When OUTP ON is executed, the power supply output is restored to the previ-
ously saved programmed values. The saved values of voltage and current can be viewed by VOLT?
and CURR? queries. .Related Commands: OUTP?. (See example, Figure C-3. Multiple channel
examples: OUTP OFF(@5,7) closes channels 5 and 7, OUTP ON(@4:7) opens channels 4, 5, 6, and
7.
C.13 OUTPut[:STATe]
QUERY
OUTP?
Syntax:
Short Form: OUTP[:STAT]?
Long Form: OUTPut[:STATe]?
Return Value:
Description: Indicates whether power supply output is enabled or disabled. Returns 0 if output disabled,
returns 1 if output enabled. Related Commands: OUTP. (See example, Figure C-3.)
C.14 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND
CURR
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
Description: Sets programmed current level at power supply output. This command programs output current to
a specific value; actual output current will depend on load conditions. If the value exceeds the maxi-
mum for the model being programmed, error message -222,”Data out of range” is posted in output
queue. (See example, Figure C-3.)
C.15 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] QUERY
CURR?
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]? MIN, MAX
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? MIN, MAX
Return Value:
Description: Returns either the programmed value, maximum value, or minimum value of current. The
CURR? query returns the programmed value of current. Actual output current will depend on load con-
ditions. The CURR?MAX query returns the maximum current allowed for a particular model. CURR?
Returns programmed current value. CURR? MAX returns maximum current allowed for power supply.
CURR? MIN returns minimum current allowed for power supply (always 0). Related Commands:
CURR. (See example, Figure C-3.)