66
8.
When the test strip has enough
control solution, the Analyzing
screen appears.
j
Wipe the tip of the bottle with a
tissue and then cap the bottle
tightly.
9.
The control result appears on
the display.
j
Select Level to display the
control solution level pop-up
menu.
7.
j
Squeeze the bottle until a tiny
drop forms at the tip.
j
Touch the drop to the
front
edge of the yellow window
of the test strip. Do not put
control solution on top of the
test strip.
52123_07006322001.indb 66