Display Shows Possible Cause(s)
Possible Solution(s)
Invalid Hypo Values
j
The hypo warning limit value
must be less than all of your
target ranges in the time
block settings.
j
Reset the hypo warning limit or revise the target
ranges in time blocks and re-enter the hypo
warning limit.
j
Enter a hypo warning limit that is below the target
ranges of your time blocks. See Chapter 5,
Changing Meter Settings.
Invalid Record
Time/Date
j
The time/date entered is
invalid (Add Data entries
cannot be set to occur in the
future).
j
Re-enter time/date.
Invalid Target
Range
j
The lower target range value
is above the upper target
range value.
j
Reset the target range values.
j
Enter the correct lower target range value and
upper target range value. See Chapter 5, Changing
Meter Settings.
j
The range selected conflicts
with your hyper and/or hypo
warning limit settings.
j
Reset range or revise warning limit settings and
re-enter range.
j
Enter the correct lower target range value and
upper target range value. See Chapter 5, Changing
Meter Settings.