Industrial Data Systems IDS 410 User Manual
Page 33
IDS 410/422 User's Manual
32
Parameter 28: TX2 Format (2 Condec data format)
If parameter 27 is set to 0 (continuous tx), parameter 28 determines what is sent:
1-AND:
status:
OL (overload), ST (stable), US (unstable, in motion)
mode:
GS (gross mode), NT (net mode)
polarity:
+ (positive weight), - (negative weight)
weight:
7 characters data, NOT zero blanked, with decimal point
units:
lb, kg, t (ton), tn (tonne), g
terminate:
lf, cr (line feed, carriage return)
2 - CONDEC
:
polarity:
space (positive weight), - (negative weight)
weight:
7 characters, zero blanked, with decimal point
units:
L (lb), K (kg), T (ton), G (gram), O (oz)
mode:
G (gross mode), N (net mode)
status:
terminate:
lf, cr (line feed, carriage return)
3 - WEIGHTRONIX
:
mode1:
space (gross mode), 0 (zero if net mode)
mode2:
G (gross mode), N (net mode)
polarity:
+ (positive weight), - (negative weight)
weight:
6 characters, NOT zero blanked, with decimal point
units:
lb, kg, oz, g
terminate:
lf, cr (line feed, carriage return)
4 - OHAS:
polarity:
space (positive weight), - (negative weight)
weight:
7 digits, zero blanked, with decimal point
units:
lb, kg, oz, g
mode:
G (gross mode), N (net mode)
status:
terminate:
lf, cr (line feed, carriage return)
5 - Label 85: Use parameter 85 (print label 85).
Print label 85: transmits the contents of parameter 85 (print label 85)
Data entered into a print label can include register data:
[253] [reg no] buffer register data, minimum size (zero blanked no spaces)
[254] [reg no] buffer data, 7 character field, zero blanked, right justified.