Parameter 5: Motion Detection Band (6)
The motion band is used by the MOTION DETECT function to determine if the scale is in
motion (not stable). If the amount of change between display updates is greater than the
motion band, a MOTION status is generated (unstable condition). The motion band is
entered in units of 1/4 dd, i.e.. For a motion band of:
1/2 display division enter 1/2 x 4 = 2.
2 display divisions enter
2 x 4 = 8.
4.5 display divisions enter
4.5 x 4 = 18.
Parameter 6: Enable Display Blanking During Motion (No Blanking)
When display blanking is enabled, the display is blanked (turned off) when the scale is in
motion (not stable). Enter a 1 (yes) to enable display blanking, enter a 0 (no) to disable
display blanking
.
Parameter 7: Zero Tracking Delay (1)
The automatic zero tracking function removes small deviations from zero. The IDS
410/422 adjusts the zero setting by approximately 1/30th of a display digit per delay
interval. The delay interval is the number of A/D conversions between adjustments.
Parameter 8: Zero tracking band (12)
The zero tracking band is the maximum displacement from zero wherein zero tracking is
active. In order for zero tracking to be active, the following conditions must be met:
• Scale weight is within Zero Track Band.
• Scale is stable (no motion)
• Zero Track Band is greater than 0 (zero).
The zero tracking band is entered in units of 1/4 dd, i.e.. for a band of:
1/2 display division enter 1/2 x 4 = 2.
2 display divisions enter 2 x 4 = 8.
4.5 display divisions enter 4.5 x 4 = 18
Parameter 9: Push Button Zero Percent (100)
Enter the percent of full-scale capacity that can be acquired as the zero point with the
ZERO key. 1 = 1%, 2 = 2%, ...... 100 = 100%.