Industrial Data Systems IDS 410 User Manual
Page 26

IDS 410/422 User's Manual
25
Parameter 1: Load Cell mv/V (3 mv/v)
The IDS 410/422 can be used with load cell outputs of 1 mv/V, 2 mv/V, or 3 mv/V. Enter
a 1, 2, or 3 for the load cell type used in your application. NOTE: there are 2 additional
settings: 0 for high gain (0.5mv/v inputs) and 4 for low gain (6 mv/v inputs).
Parameter 2: Digital Filter (9)
The digital filter removes scale vibrations and other noise signals from the weight display.
The filter range is between 0 and 18. A setting of 0 provides no filtering; a setting of 18
provides maximum filtering. The higher the filters number the longer the settle time of the
weight display. For example:
Filter Setting
Settle Time
3
0.4 sec
6
0.8 sec
9
1.75 sec
12
3.5 sec
15
7.0 sec
Use the smallest filter setting that provides satisfactory display stability
.
Parameter 3: Display Update Rate (0)
Enter the time between display updates (in tenths of a second). An update rate of 1 causes a
0.1 second delay between display updates, a rate of 200 causes a 20 second delay between
updates. NOTE: motion detection is calculated using the weight values of the display rate.
If the update rate is 50, the motion calculation is based on weight readings sampled at
0.5sec intervals.
Parameter 4: Motion Detection Delay (15)
Motion detection delay is used by the motion detection function to determine if the scale is
in motion (not stable). Enter the amount of time (in tenths of a second) that the scale must
be stable for motion status to be false (scale status = stable).
Example:
0 = disables motion detection.
2 = 0.2 second delay before motion = false (stable = true).
15 = 1.5 second delay.
30 = 3.0 second delay.