beautypg.com

Hardy ANY-WEIGH Bench Scales User Manual

Page 25

background image

Chapter 5 - Setup

5-3

Parameter 21 = Calibrate High Command
0 to 1 (Set to True to Complete Command)
Length in Bytes = 1

Parameter 22 = Calibrate using C2

®

0 to 1 (Set to True to Complete Command)
Length in Bytes = 1

Parameter 23 = Span Weight in Kgs
0.000 to 2147483.647 0.045
Length in Bytes = 4

Parameter 30 = Net Weight in Kgs
Read Only
Length in Bytes = 4

Parameter 31 = Gross Weight in Kgs
Read Only
Length in Bytes = 4

Parameter 32 = Tare Weight in Kgs
-999999 - 999999 0.00
Length in Bytes = 4

Parameter 34 = A/D Counts
Length in Bytes = 4

Parameter 35 = Calibration Low Weight in Lbs.
Length in Bytes = 4

Parameter 36 = Calibration Low Weight in Kgs.
Length in Bytes = 4

The “Calibration Low Weight” parameter speci-
fies the weight on the scale when the low step of
a calibration is done in Traditional Calibration
and is the Reference Point for C2 Calibration.

Parameter 37 = Weight Multiplier
Length in Bytes = 4

This integer parameter can be set to 1, 10, 100,
etc. to allow the user to select the number of dec-