Hardy ANY-WEIGH Bench Scales User Manual
Page 25
Chapter 5 - Setup
5-3
•
Parameter 21 = Calibrate High Command
0 to 1 (Set to True to Complete Command)
Length in Bytes = 1
•
Parameter 22 = Calibrate using C2
®
0 to 1 (Set to True to Complete Command)
Length in Bytes = 1
•
Parameter 23 = Span Weight in Kgs
0.000 to 2147483.647 0.045
Length in Bytes = 4
•
Parameter 30 = Net Weight in Kgs
Read Only
Length in Bytes = 4
•
Parameter 31 = Gross Weight in Kgs
Read Only
Length in Bytes = 4
•
Parameter 32 = Tare Weight in Kgs
-999999 - 999999 0.00
Length in Bytes = 4
•
Parameter 34 = A/D Counts
Length in Bytes = 4
•
Parameter 35 = Calibration Low Weight in Lbs.
Length in Bytes = 4
•
Parameter 36 = Calibration Low Weight in Kgs.
Length in Bytes = 4
The “Calibration Low Weight” parameter speci-
fies the weight on the scale when the low step of
a calibration is done in Traditional Calibration
and is the Reference Point for C2 Calibration.
•
Parameter 37 = Weight Multiplier
Length in Bytes = 4
This integer parameter can be set to 1, 10, 100,
etc. to allow the user to select the number of dec-