Universal scale electrical connections, Universal scale electrical connections -3, Fig. 3-3 leveling scale -3 – Hardy ANY-WEIGH Bench Scales User Manual
Page 15

Chapter 3 - Installation
3-3
FIG. 3-3 LEVELING SCALE
Step 4.
Rotate the spirit level 90° and check for
level again.
Step 5.
Adjust the leveling feet if necessary.
Universal Scale
Electrical Connections
Make sure the cable is not pinched and is clear of the
feet. The Bench Scale requires a recommended DC or
AC excitation voltage of 5 volts with a maximum of
15 volts. Wire the Scale cable to an instrument or
indicator following the color code label found on the
upper frame or as follows:
+ Excitation
Red
+ Sense
Blue
+ Signal
Green
+ C2 Cal
Gray
- Signal
White
- Sense
Brown
- Excitation
Black
- C2 Cal
Violet
WARNING
USE CAUTION WHEN WIRING. MAKE SURE
TO DISCONNECT ALL POWER TO THE
SCALE. FAILURE TO REMOVE POWER CAN
RESULT IN SCALE OR INSTRUMENT DAM-
Spirit Level