7 maintenance, 1 replacement of the device fuse, Maintenance 7.1 replacement of the device fuse – FRITSCH LABORETTE 24 with V-shaped channel User Manual
Page 26: 7maintenance

7
Maintenance
Apart from cleaning, the LABORETTE 24 vibratory feeder is maintenance
free.
DANGER!
Mains voltage
– Before beginning with maintenance work, unplug the
mains plug and protect the device against being unin-
tentionally switched back on again!
– Indicate maintenance work with warning signs.
– Maintenance work may only be performed by special-
ised personnel.
– Put safety equipment back into operation after mainte-
nance or repair work.
The most important element of maintenance is regular
cleaning:
Observe the regulations of the trade association (BGV A3)
when cleaning the entire device ‐ especially if the device is
set up in a dusty environment or is used to process dusty
source material.
We recommend keeping a safety logbook Ä Chapter 12
‘Safety logbook’ on page 34, where all work (maintenance,
repairs......) performed on the device is entered.
7.1 Replacement of the device fuse
The LABORETTE 24 is protected by a T1.25A/250V glass fuse.
Pull the mains plug out of the socket.
DANGER!
Danger to life due to power surges
Exposed power contacts
You could touch the fuses or fuse holder with live contacts
when changing the fuses. Electric shocks can cause burns
and cardiac arrhythmia or respiratory and cardiac arrests.
Remove the mains cable before changing the fuses.
Maintenance
- 26 -