beautypg.com

Heading1 - 2.5 setting parameter units, Heading1 - 2.8 configuring rdo pro options, 5 setting parameter units – In-Situ Con TROLL PRO Operators Manual (purchased before 10 May 2012) User Manual

Page 12: 7 setting the sentinel value (offline/error value), 8 configuring rdo pro options, 1 entering barometric correction values

background image

12

Configuring Probes

1. Select Menu, Settings, Probes.
2. Select a probe, then select parameters.
3. Select each parameter from the list and select Disable to stop logging or viewing

that parameter.

2.5

SETTING PARAMETER UNITS

1. Select Menu, Settings, Probes.
2. Select a probe.
3. Select parameters.
4. Use the arrow keys to scroll to the selected parameter. Select Enter.
5. Select Units.
6. Use the arrow keys to scroll to the correct unit of measure. Select Enter.

2.6

SETTING PARAMETER RESOLUTION (SETTING THE DECIMAL PLACE)

1. Select Menu, Settings, Probes.
2. Select a probe. (Remember that Probe C is the controller.)
3. Select parameters.
4. Use the arrow keys to scroll to the selected parameter. Select Enter.
5. Select Resolution.
6. Use the arrow keys to set the resolution. Select Enter.

2.7

SETTING THE SENTINEL VALUE (OFFLINE/ERROR VALUE)

To set a value to display when the probe returns an error condition for a parameter:

1. Select Menu, Settings, Probes.
2. Select a probe. (Remember that Probe C is the controller.)
3. Select parameters.
4. Use the arrow keys to scroll to the selected parameter. Select Enter.
5. Select Sentinel.
6. Use the arrow keys to set the sentinel (error) value. Select Enter.

2.8

CONFIGURING RDO PRO OPTIONS

2.8.1

Entering Barometric Correction Values

1. Select Menu, Settings, Probes.
2. Select the RDO PRO probe.
3. Select Options, then Barometer.
4. Choose between Fixed or Automatic compensation.

• If you select “Fixed”, you must then enter a barometric value.
• If you select “Automatic”, the RDO probe will pull a barometric value from the

controller.

Table 3. Sensor parameters that can be enabled or disabled

RDO PRO

Level TROLL

Aqua TROLL

Con TROLL PRO

Concentration

Pressure

Pressure (Aqua TROLL 200 only)

Barometer

Temperature

Temperature

Temperature

Saturation

Depth/Level

Depth/Level
(Aqua TROLL 200 only)

Temperature

Pressure

Actual Conductivity

Specific Conductivity

Salinity

TDS

Resistivity

Density

This manual is related to the following products: