HT instruments MACROTEST 5035 User Manual
Page 45
MACROTEST
5035
EN - 44
If the instrument detects that the earth
cable (green) is not connected, the
screen alongside is displayed for 5
seconds then the initial display is back.
Check the connections of PE conductor
under test.
RCD LOOP
o
>
1999
Ω
no
PE
If a contact voltage
U
t
higher than the
selected limit (U
L
) is detected the
instrument
interrupts the test and emits
a long sound signal at the end of the
test and displays the values
alongside.displays the screen
alongside.
RCD LOOP
o
>
50
V
A
UL
If no power is detected during the test
the instrument displays the screen
alongside.
RCD
LOOP
o c
- - - - - -
If the thermistores of the instrument are
damaged, it displays the screen
alongside.
RCD
LOOP
o
c
If the instrument gets overheated it
does not permit to effect the tests and
displays the message alongside. Wait
until the instrument displays the initial
screen before proceeding with
measuring.
RCD
LOOP
o
THE PREVIOUS RESULTS CANNOT BE SAVED.
Message
"no PE": the
instrument does
not detect the
protection circuit.
SAVE
Message
"Att UL": a too
high contact
voltage was
detected
Message
">50" o “>25V”:
the instrument detects a contact voltage higher
than the selected limit (50V in this case)
Messagge
“hot”:
the instrument is
overheated.