O 5.75, O> 99.9 – HT instruments MACROTEST 5035 User Manual
Page 14

MACROTEST
5035
EN - 13
4.1.3.1. Mode
"AUTO"
At the end of the test, if the
average
resistance value Ravg results to be
lower than 5
Ω the instrument emits a
double sound signal indicating the
positive outcome of the test and
displays one screen similar to the
screen alongside.
CAL
LOW
Ω
1.07
Ω
219
mA
AUTO
SAVING:
The test can be stored pressing the
SAVE key twice
(according to paragraph 5.1).
4.1.3.2. Mode "R+TIMER" or “R-TIMER"
If during the test a
resistance value
R+Timer or R–Timer lower than 5
Ω
was detected, the instrument (after the
set time has elapsed) emits a
double
sound signal indicating the positive
outcome of the test displays one screen
similar to the screen alongside.
CAL LOW
Ω
1.08
Ω
219
mA
5
s
R+
R-
TIMER
SAVING:
The test can be stored pressing the
SAVE key twice
(according to paragraph 5.1).
4.1.4. Anomalous cases during "AUTO", "R+TIMER", "R-TIMER" tests
In case a value of
Ravg or R+ or R-
higher than or equal to 5
Ω but lower
than 99.9
Ω was detected, at the end of
the test the instrument
emits a long
sound signal and displays one screen
similar to the screen alongside.
CAL LOW
Ω
o 5.75
Ω
216
mA
5
s
AUTO
In AUTO mode, if a
Ravg or R+ or R-
higher than 99.9
Ω was detected, at
the end of the test the instrument emits
a long sound signal and displays the
screen alongside.
CAL LOW
Ω
o> 99.9
Ω
- - -
mA
AUTO
Average test
current value
Iavg.
Average
resistance value
Ravg.
Value of the test
current
I+ or I-.
ATTENTION: value of R
avg
higher than 5
Ω
99.9
Ω
is the
maximum value
which can be
measured in the
LOW
Ω AUTO or
R+ or R-mode.
The symbols
R
+
or
R
-
are displayed.
AUTO mode.
AUTO mode.
ATTENTION: value of R
avg
or R+ or R-
is too
high.
Duration time of
the test.
Only in case the
mode R+TIMER
or R-TIMER was
selected.
SAVE
SAVE