Perma Pure GASS-2040 User Manual
Page 18

18
GASS-2040 User Manual | Original Instructions
4.7. Electrically Heated Systems
1.
Check that no condensate is present in sample line between filter and dryer.
2.
If water droplets are visible, increase temperature of filter/dryer compartment by
5
°C.
3.
Allow about ½ hour for system to stabilize and then check again for condensate.
4.
If necessary, continue to increase temperature at 5
°C intervals until condensate
is no longer visible. Do not exceed 100
°C! Maximum system temperature is
100
°C. Operation above this temperature can cause damage to Nafion dryer.
It is essential that purge gas flows continuously and setpoint temperature is not
above 100°C to prevent damage to dryer.