b) for output to the serial interface (RS232), the receiving device (normally a PC) is set to the data

transfer parameters of the Rheometer (see chapter 6.7.5 “RS232 Parameters”) and ready to receive
data. If the receiving side is not ready, the data will not be transmitted or will be transmitted to Null.
(See also chapter 6.8 “Serial data transfer”).

VI.7.2 Configuration

MeasCount mode

Measurement series counter MEASCOUNT is a counter, increasing its count by 1 when starting a
measurement. Measurement series counter serves, as well as ID of the measurement for identification
of a measurement series. Counters’ count rises until automatic of manual reset (“Reset meascount”) will
be performed.

Using the function “Set meascount-mode” you set the counter mode (if the measurement series counter
will be daily reset or not).








After choosing one of two reset modes you again come to the Configuration-Menue.

VI.7.3 Configuration

Reset meascount


This function resets the measurement series counter (regardless a MEASCOUNT mode set) = manual
reset of the measurement series counter.

No further selections or settings are required.

VI.7.4 Configuration

Set Time/Date

This function allows the user to set the internal clock and internal calendar.

Setting of time is carried out in 24 hour format.






Setting of date is carried out as dd.mm.yy. :





Select meascount mode:
- No reset
- Reset every day

Set Time

Hour: 17
Minutes: 15

Set Date

Day: 17
Month: 03
Year: 01

Reset only via function
„Reset meascount“

Automatic resetting every
new day

Minutes 00 to 59

Month 01 to 12

Hours 00 to 23

Day 01 to 31

Year 00 to 99