VI. Operation and Menue System


In the following chapter you will get overview on operations, menue system of R/S Rheometer and
measurements, both in manual and PC controlled modes.

The menue-system of R/S Rheometer (firmware-version 4.02) is a dual-language system. Following
languages are software selectable:

• English
• German

Desired language can be selected via menue item Configuration

Language (if English language is

active) or Konfiguration

Sprache (if German language is active).

More details on language selection see chapter 6.7.6.
The following chapter will describe instrument handling with ENGLISH language active.

After switching on the R/S Rheometer, LCD indicates the following information for few seconds:

* Name of the Rheometer;
* Software firmware-version installed in the instrument (Brand name);
* Serial number of the instrument (xxxxxx);
* Date and time;

Example:





After about 5 sec the instrument checks voltages of the Power Supply Unit. Displayed voltages are to lie
in the range from 14.9 to 16 V. Otherwise, some technical fault has occurred (

→ Repairs).


Beispiel:




This message is displayed for about 3 seconds. Then the MAIN menue is displayed.
Before you will get overview of the menue system of R/S Rheometer (Chapter 6.4), at this place it is
worth to make a brief summary on the rheometer keyboard and corresponding rheometer functions.

R/S Rheometer

Ver.: 4.02 #xxxxxx
07.05.98 15.12
© Brookfield

Voltage check:
VCC:15.25 V