Analog Input and Output Calibration Procedures

C-2

4.

Connect the UDAS to the host computer and allow at least a 15 minute warm-up period before

performing the adjustments.

FIGURE C.1 Locations of Potentiometers for Adjusting Analog Circuits on page C-2 shows the

locations of the potentiometers used to adjust the analog circuits on a UDAS unit.

FIGURE C.1 Locations of Potentiometers for Adjusting Analog Circuits

CAUTION:

The unit may become damaged if you use an uninsulated screwdriver
and accidentally touch other components.