F 40
Format switch cassette incorrect.
Check 4 format switches.
F 41
M1 motor breaker stuck/defective.
Check K1 breaker.
F 42
M2 motor breaker stuck /defective.
Check K2 breaker.
F 43
M3 motor breaker stuck /defective.
Check K3 breaker.
F 44
M4 motor breaker stuck /defective.
Check K4 breaker.
F 45
M5 motor breaker stuck /defective.
Check K5 breaker.
F 46
M6 motor relay stuck /defective.
Check K19 relay (CPU board).
F 47
FU0 N>0 active.
FU0 relay defective/ input
defective / check cabling.
F 48
FU1 N>0 active.
FU1 relay defective / input
defective / check cabling.
F 50
Collimator malfunction.
Check cabling, potentiometers,
setup parameters.
F 61
Limit switch, temperature, breaker
malfunction, M1 tilt /vertical drive or M3.
Free run (press button on breaker
board). Allow motor to cool down.
F 62
Limit switch, temperature, breaker
malfunction, M2 system drive.
Free run (press button on breaker
board). Allow motor to cool down.
F 64
Limit switch, temperature, breaker
malfunction, M4 table longitudinal drive.
Free run (press button on breaker
board). Allow motor to cool down.
F 79
Invalid parameter version
Initialize correct parameter
version.
F 80
EEPROM memory error.
Download new parameter file.
F 99
Unit is in the service mode.
Set the service switch to the RUN
mode.
F = Fatal Error = System error = cannot be corrected