4.2 Function Check


Switch on the system. Initiate movement through the entire movement path.



Vertical drive:

- Even movement?
- Automatic stop in the bottom and top positions?

- Soft start and smooth movement out?
- Movement initiated from footswitch?

- Movement initiated from manual control unit?
- Any uninitiated lowering with load applied?




Tilt drive:

- Even movement?
- Soft start and smooth movement out?

- Automatic stop in vertical limit position (88°) ?
- Automatic stop in horizontal limit position (0°) ?
- Automatic stop in Trendelenburg position (-20°) ?
- Movement initiated from footswitch?
- Movement initiated from manual control unit?

- Any uninitiated lowering with load applied?



System longitudinal movement:

- Even movement?
- Soft start and smooth movement out?

- Automatic stop in head-end limit position?
- Automatic stop in foot-end limit position?
- Movement initiated from footswitch?
- Movement initiated from manual control unit?
- Any movement with opened Bucky flap?





Tabletop longitudinal movement:

- Even movement?
- Soft bootup and soft outward movement?

- Automatic stop in the middle position?
- Automatic stop in the head-end limit position?
- Automatic stop in the foot-end limit position?

- Movement initiated from footswitch?

- Movement initiated from manual control unit?