Tabletop transverse movement:

- Even movement?
- Soft startup and soft outward movement?
- Automatic stop in the centered position?
- Automatic stop in the limit position?

- Movement initiated from footswitch?

- Movement initiated from the manual control unit?




Movable Tube Unit Support Arm:

- Easy separation in the exposure position?
- Easy separation in the park position?

- Even movement?

- Positive engagement into the limit positions?


Table extension:

- Does the table extension engage in place in the opening at the foot end?
- Is the table extension blocked from sliding out?
- Does the controller when the table extension is controlled?



Footrest:

- Does the footrest securely and easily engage the footrest?


Micturation seat:

- Does the micturation seat engage in place in the opening?

- Is the micturation seat secured against falling out?
- Does the controller sense the micturation seat?


Accessories:

- Are all configured accessories functional?




General:

- Is the centering of the tube unit to the adjustable Bucky okay?

- Is the centering of the tube unit to the I.I. okay?

- Is radiation blocked when the tube unit support arm is in the park position?