Hansol Multitech Uromat 3000 User Manual
Page 84
0116 7201 -84 of 90 - 06/00
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
5. Troubleshooting
Error
Number
Cause
Corrective Measure
F 1
Frequency converter 0 for tilt/vertical,
system failed.
Check connection cables, settings
(address, Baud rate).
F 2
Frequency converter 1 for tilt/vertical, table
longitudinal movement, cassette failed.
Check connection cables, settings
(address, Baud rate).
F 3
Frequency converters 0+1 failed.
Check connection cables, settings
(address, Baud rate).
F 10
Footswitch communication problem.
Check fuse, cabling.
F 11
Manual control unit communication
problem.
Check fuse, cabling.
F 13
Incorrect footswitch software version.
Replace footswitch.
F 14
Incorrect manual control unit software
version.
Replace manual control unit.
F 20
Analog voltage failed.
Check fuse.
F 21
M1 tilt/vertical potentiometer error.
Check potentiometer cables and
potentiometer.
F 22
System potentiometer error.
Check potentiometer cables and
potentiometer.
F 23
M3 tilt/vertical potentiometer error.
Check potentiometer cables and
potentiometer.
F 24
Table longitudinal potentiometer error.
Check potentiometer cables and
potentiometer.
F 25
Cassette potentiometer error.
Check potentiometer cables and
potentiometer.
F26
Table transverse potentiometer error /
motor defective.
Check motor potentiometer cables
and potentiometer.
F 28
Collimator vertical potentiometer error.
Check potentiometer cables and
potentiometer.
F 30
Button fuse malfunction.
Check buttons, cables.
F 31
Footswitch button sticking.
Check buttons.
F 32
Manual control unit button sticking.
Check buttons.
F 33
Exposure switch active during system start. Check switches.
F 34
Fluoroscopy switch active during system
start.
Check switches.
F 35
Exposure centering switch defective.
Check switches, cabling.