Amprobe BDM40-UA Bench-Digital-Multimeter User Manual
Page 27

26
DC CURRENT:
RANGE
RESOLUTION
ACCURACY for 1 Year
VOLTAGE BURDEN
200µA
2mA
20mA
200mA
0.01 µA
0.1 µA
1 µA
10 µA
(0.2% of reading + 2 digits)
1 mV / 1
μA
100 mV / 1 mA
10 mV / 1 mA
1 mV / 1 mA
200 mV / 1 A
2000mA
20A *
100 µA
1 mA
(0.3% of reading + 2 digits)
14 mV / 1 A
* Overload Protection: 5 ranges (200µA, 2mA, 20mA, 200mA, 2000mA), with fuse protection,
20A range, no fuse.
AC CURRENT (True RMS Responding, AC or AC+DC):
INPUT
CURRENT
RESO-
LUTION
Range
20Hz**
45Hz
2kHz
10kHz -
20kHz
VOLTAGE
BURDEN
10µA -200µA
.01µA
200µA
1 mV / 1
μA
100µA -2mA
0.1 µA
2mA
100mV / 1 mA
1mA-20mA
1 µA
20mA
1%
0.5%
1%
2%
10 mV / 1 mA
10mA-200mA 10
µA 200mA
+15
+15 +15 +15
1mV
/
1
mA
100mA-2000mA
100
μA
2000m
A
200 mV / 1 A
2000mA-20A 1
mA 20
A
Not Specified
14mV / 1 A
* * Typically 3 to 5 digits of drift will be observed at full scale at 20Hz.
Crest Factor Range: Waveforms with a Peak / RMS ratio of 1:1 to 3:1 at full scale.
RESISTANCE
RANGE
RESOLUTION
ACCURACY for 1 Year
FULL SCALE VOLTAGE ACROSS
UNKNOWN RESISTANCE
00
Ω 0.01Ω
± (0.1% reading + 4 digits)
0.2V
2k
Ω
20k
Ω
200k
Ω
0.1
Ω
1
Ω
Ω
10
Ω
± (0.1% reading + 4 digits)
2V
2V
0.2V
2000k
Ω
20M
Ω
100
Ω
1k
Ω
± (0.25% reading + 4 digits)
2V
2V
Overload Protection: 250V DC/AC rms on all ranges.