BOP-VISA 010906

1.2.5

PATTERN GENERATION, EXAMPLE 2

To generate a 100 point voltage complex waveform comprised of 1/4 cycle sine wave and 1/4 cycle trian-
gle wave with an amplitude of 10 volt peak to peak riding on 0 volt level, and a total duration of 1 second,
enter the parameters listed in Table 2.

By using combinations of the four basic waveform types, complex waveforms can be generated, and
repeated using the Pattern Execution window (see PAR. 1.2.6) to provide great flexibility in determining the
output.

1.2.6

PATTERN EXECUTION

The Pattern Execution window (Figure 11) allows the program established by the pattern generation win-
dow to be executed. If the Loop Direction is set to UP, the Loop From window establishes the point in the
loop that will start repeating. For example, if the list is 10 points, and Loop From is set to 3, the first 2
points would only be executed once, and the points from 3 to 10 would be repeated the number of times
specified in the Loop Count Window. If Loop Count is set to 0 the program will run continuously until
stopped by the user. The RUN button starts the program. While the program is running the Running indi-
cator is on and the RUN button changes to a STOP button used to halt the program. When the program is
halted, the unit reverts to the settings in effect prior to issuing the RUN command.

TABLE 1. PATTERN GENERATION, EXAMPLE 1

PARAMETER

ENTER

RESULT

Destination

VOLTAGE

Press GENERATE button after all parameters entered to
add 100 points to list which will produce the following out-
put:

Waveform

SINE

Frequency (Hz)

1.000

Start Angle

0.000

End Angle

360.000

Amplitude (p-p)

1.000

Waveform Offset

28.000

Current

(blank)

Points

100